Related Experiment Video
Updated: Feb 25, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Radial MRI with variable echo times: reducing the orientation dependency of susceptibility artifacts of an MR-safe
Katharina E Schleicher1, Michael Bock2, Klaus Düring3
1Department of Radiology, Medical Physics, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Breisacher Strasse 60a, 79106, Freiburg, Germany. katharina.schleicher@posteo.de.
Objectives:
Guidewires are indispensable tools for intravascular MR-guided interventions. Recently, an MR-safe guidewire made from a glass-fiber/epoxy compound material with embedded iron particles was developed. The size of the induced susceptibility artifact, and thus the guidewire's visibility, depends on its orientation against B 0. We present a radial acquisition scheme with variable echo times that aims to reduce the artifact's orientation dependency.
Materials And Methods:
The radial acquisition scheme uses sine-squared modulated echo times depending on the physical direction of the spoke to balance the susceptibility artifact of the guidewire. The acquisition scheme was studied in simulations based on dipole fields and in phantom experiments for different orientations of the guidewire against B 0. The simulated and measured artifact widths were quantitatively compared.
Results:
Compared to acquisitions with non-variable echo times, the proposed acquisition scheme shows a reduced angular variability. For the two main orientations (i.e., parallel and perpendicular to B 0), the ratio of the artifact widths was reduced from about 2.2 (perpendicular vs. parallel) to about 1.2 with the variable echo time approach.
Conclusion:
The reduction of the orientation dependency of the guidewire's artifact via sine-squared varying echo times could be verified in simulations and measurements. The more balanced artifact allows for a better overall visibility of the guidewire.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...