Evaluation of a new free light chain ELISA assay: bringing coherence with electrophoretic methods
Joannes F M Jacobs1, Corrie M de Kat Angelino2, Huberdina M L M Brouwers2
1Radboud University Medical Center, Department of Laboratory Medicine; Laboratory of Medical Immunology (Route 469), Geert Grooteplein 10, 6525 GA Nijmegen, The Netherlands, Phone: +31 (0)24-3617414, Fax: +31 (0)24-3619415.
Background:
Serum free light chain (sFLC) measurements are increasingly important in the context of screening for monoclonal gammopathies, prognostic stratification, and monitoring of therapy responses. At the same time, analytical limitations have been reported with the currently available nephelometric and turbidimetric sFLC assays. We have evaluated a new quantitative sFLC ELISA for its suitability in routine clinical use.
Methods:
Reference ranges of the Sebia FLC assay were calculated from 208 controls. Assay interference, reproducibility, lot-to-lot variability, and linearity were assessed. Method comparison to the Freelite assay (Binding Site) was conducted by retrospective analysis of 501 patient sera.
Results:
Reference ranges of the Sebia κ/λFLC-ratio were 0.37-1.44. We observed good sensitivity (1.5 mg/L) and linearity in both polyclonal and monoclonal sFLC samples and never experienced antigen excess. Sebia FLC reproducibility varied between 6.7% and 8.1% with good lot-to-lot consistency. Method comparison with Freelite showed the following correlations: κFLC R=0.94, λFLC R=0.92 and κ/λFLC-ratio R=0.96. The clinical concordance of the κ/λFLC-ratio of both methods was 94%. Significant quantitative differences were observed between both methods, mainly in sera with high FLC concentrations. The Sebia monoclonal FLC concentrations were coherent with those obtained by serum protein electrophoresis (SPE). Freelite monoclonal FLC concentrations were consistently higher, with a mean 12-fold overestimation compared to SPE.
Conclusions:
The Sebia FLC assay provides a novel platform for sensitive and accurate sFLC measurements. The Sebia FLC showed good clinical concordance with Freelite. Further studies are warranted to confirm the clinical value of this assay.
More Related Videos
09:32Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method
Published on: September 10, 2017
11:03Use of Capillary Electrophoresis Immunoassay to Search for Potential Biomarkers of Amyotrophic Lateral Sclerosis in Human Platelets
Published on: February 10, 2020
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...