The effect of scatter correction on planar and tomographic semiquantitative 123I cardiac imaging. A phantom study
Emmanouil Papanastasiou1, Efstratios Moralidis, Anastasios Siountas
1Department of Medical Physics, AHEPA University Hospital, Aristotle University Medical School, Thessaloniki, Greece. empapana@med.auth.gr.
Objective:
In cardiac I-123 (123I) imaging downscatter from high energy emissions degrades the image and introduces distortion of semi-quantitative analysis when using a low energy collimator. The effect of a triple energy window (TEW) scatter correction technique, using windows immediately above and below the principal window centered on 159keV, was examined.
Materials And Methods:
A hemispherical cardiac phantom was inserted into a cylindrical phantom and both were filled with radioactive 123I water solutions. Phantoms were submitted to planar and tomographic scintigraphy under various acquisition and processing conditions, including the use of medium energy (ME) and low energy (LE) collimation.
Results:
In planar imaging, there was a distance dependent count loss with the LEHR collimator which was partly restored with TEW correction. There was minimal dependence of count rate with distance in using ME collimation. Conversely, the heart to background (H/B) ratio increased with increasing distance with the LEHR collimator, but in applying the TEW correction that ratio paralleled the minimally affected values obtained with the ME collimation. In tomographic imaging the acquired H/B ratio was lower with LE collimation alone, in comparison to the ME collimator, but it was raised significantly when applying the TEW scatter correction. Quantitative measurements also depended on the background method and the reconstruction algorithm applied.
Conclusion:
In cardiac 123I imaging with a LE collimator the use of TEW scatter correction provides a semi-quantitative assessment comparable to that attained with ME collimation and may moderate inter-institutional inconsistencies.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography