Performance of life support breathing apparatus for under-ice diving operations
Michael A Lang1, John R Clarke2
1University of California San Diego, Department of Emergency Medicine, San Diego, California, U.S.
Introduction:
Single-hose scuba regulators dived in very cold water may suffer first- or second-stage malfunction, yielding complete occlusion of air flow or massive freeflow that rapidly expends a diver's air supply.
Purpose:
This study, conducted in Antarctica, evaluated the under-ice performance of a sampling of commercially available regulators.
Methods:
Seventeen science divers logged a total of 305 dives in -1.86°C seawater under 6-meter-thick Antarctic fast-ice over two field seasons in 2008 and 2009. Dive profiles had an average depth of 30 msw and dive time of 29 minutes, including a mandatory three-minute safety stop at 6 msw. Sixty-nine unmodified regulator units (17 models) from 12 different manufacturers underwent standardized pre-dive regulator care and were randomly assigned to divers. Depths and times of onset of second-stage regulator freeflow were recorded.
Results:
In 305 dives, there were 65 freeflows. The freeflows were not evenly distributed across the regulator brands. Regulator failure rates fell into two categories (⟨ 11% and ⟩ 26%). The regulators classified for the purpose of the test as "acceptable" (⟨ 11% failure rate: Dive-Rite Jetstream, Sherwood Maximus SRB3600, Poseidon Xstream Deep, Poseidon Jetstream, Sherwood Maximus SRB7600, Poseidon Cyklon, Mares USN22 Abyss) experienced only nine freeflows out of 146 exposures for a 6% overall freeflow incidence. Those classified as "unacceptable" (⟨ 26% failure rate) suffered 56 freeflows out of 159 exposures (35% freeflow incidence.).
Conclusions:
Contrary to expectations, the pooled incidences for the seven best performing regulators was significantly different by Chi-square test from the 10 remaining regulators (P ⟨ 0.001).
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...