Comparison of stroke volume measurements during hemodialysis using bioimpedance cardiography and echocardiography
Michael J Germain1, Jyovani Joubert2, Daniel O'Grady3
1Baystate Medical Center, Springfield, Massachusetts and University of Massachusetts Medical School, Worcester, Massachusetts, USA.
Background:
Fluid management remains a major challenge of hemodialysis (HD) care, with serious implications for morbidity and mortality. Intradialytic fluid management is typically guided by blood pressure, an indirect resultant of hemodynamics status. Direct measurements of hemodynamic parameters may improve cardiovascular outcomes by providing rational bases for intervention. We compare stroke volume (SV) measurements using a noninvasive, regional biompedance cardiography device (NiCaS) with Doppler echocardiography (Echo) in HD setting.
Methods:
Stroke volumes were simultaneously measured using the devices in 17 patients receiving maintenance HD. Measurements were made during 2 weekly HD treatments, and twice within each HD treatment during the first and last hour of each treatment, for a total of 64 SV measurements. Agreement between devices was assessed using linear regression, a Pearson's correlation coefficient, and a Bland-Altman plot all adjusted for repeated measures within patients.
Results:
Echo and NiCaS SV mean and 95% CIs were 58.0 (50.1, 65.8) and 56.7 (49.4, 64.0) mL, respectively. NiCaS SV correlated strongly with Echo SV during the first and last hours of treatments (r = 0.93, P < 0.001 and r = 0.92, P < 0.001, respectively). Linear regression of NiCaS on Echo showed a slope of 0.97, 95% CI (0.91, 1.02) which did not differ from 1, P = 0.20. A Bland-Altman plot and 4-Quadrant plot confirmed that the 2 methods produced comparable measurements.
Conclusion:
NiCaS SV measurements are similar to and strongly correlated with Echo SV measurements. This suggests that noninvasive NiCaS technology may be a practical method for measuring SV during HD.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...