Dark-Adapted Chromatic Perimetry for Measuring Rod Visual Fields in Patients with Retinitis Pigmentosa
Lea D Bennett1, Martin Klein1, Kirsten G Locke1
1Retina Foundation of the Southwest, Dallas, TX, USA.
Purpose:
Although rod photoreceptors are initially affected in retinitis pigmentosa (RP), the full-field of rod vision is not routinely characterized due to the unavailability of commercial devices detecting rod sensitivity. The purpose of this study was to quantify rod-mediated vision in the peripheral field from patients with RP using a new commercially available perimeter.
Methods:
Participants had one eye dilated and dark-adapted for 45 minutes. A dark-adapted chromatic (DAC) perimeter tested 80 loci 144° horizontally and 72° vertically with cyan stimuli. The number of rod-mediated loci (RML) were analyzed based on normal cone sensitivity (method 1) and associated with full-field electroretinography (ERG) responses by Pearson's r correlation and linear regression. In a second cohort of patients with RP, RML were identified by two-color perimetry (cyan and red; method 2). The two methods for ascribing rod function were compared by Bland-Altman analysis.
Results:
Method 1 RML were correlated with responses to the 0.01 cd.s/m2 flash (P < 0.001), while total sensitivity to the cyan stimulus showed correlation with responses to the 3.0 cd.s/m2 flash (P < 0.0001). Method 2 detected a mean of 10 additional RML compared to method 1.
Conclusions:
Scotopic fields measured with the DAC detected rod sensitivity across the full visual field, even in some patients who had nondetectable rod ERGs. Two-color perimetry is warranted when sensitivity to the cyan stimulus is reduced to ≤20 dB to get a true estimation of rod function.
Translational Relevance:
Many genetic forms of retinitis pigmentosa (RP) are caused by mutations in rod-specific genes. However, treatment trials for patients with RP have relied primarily on photopic (cone-mediated) tests as outcome measures because there are a limited number of available testing methods designed to evaluate rod function. Thus, efficient methods for quantifying rod-mediated vision are needed for the rapidly increasing numbers of clinical trials.