A Murine Model of Irreversible and Reversible Unilateral Ureteric Obstruction
Published on: December 20, 2014
Extraction of trapped double J in ureteroscope: a novel technique
1Urology Department, Baharloo Hospital, Tehran University of Medical Sciences, Tehran - Iran.
Introduction:
Urological endoscopes and percutaneous devices are expensive and their maintenance and cleaning need special attention. Sometimes urologists can help repairing their devices.
Case Description:
During taking out a Double J, Double J was trapped in the working channel of ureteroscope unintentionally. It was not possible to take it out by pushing methods. Finally, Double J was pulled out by a changed nephrostomy needle to a small curve hook.
Conclusions:
Using the hand-made hook to pull out a trapped Double J from ureteroscope does not destroy the device.