Objective:
Implantable cardioverter defibrillators (ICDs) are subject to electromagnetic interference (EMI). The aim of this study was to assess both the EMI of dental equipments with ICDs and related factors.
Materials And Methods:
High- and low-speed handpieces, an electric toothbrush, an implant motor and two types of ultrasonic devices were tested next to an ICD with different sensitivity settings. The ICD was immersed in a saline solution with electrical resistance of 400-800 ohms to simulate the resistance of the human body. The dental equipments were tested in both horizontal (0°) and vertical (90°) positions in relation to the components of the ICD. The tests were performed with a container containing saline solution, which was placed on a dental chair in order to assess the cumulative effect of electromagnetic fields.
Results:
The dental chair, high- and low-speed handpieces, electric toothbrush, implant motor and ultrasonic devices caused no EMI with the ICD, irrespective of the program set-up or positioning. No cumulative effect of electromagnetic fields was verified.
Conclusions:
The results of this study suggest that the devices tested are safe for use in patients with an ICD.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
11:28Concurrent EEG and Functional MRI Recording and Integration Analysis for Dynamic Cortical Activity Imaging
Published on: June 30, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Cardiopulmonary Resuscitation III: AED Use
Mechanism of Cardiac Arrhythmias