Real-Time Ultrasound Doppler Enhances Precision in Image-Guided Approaches to the Cerebellopontine Angle
Davit Baghdasaryan1, Marcel Albrecht2, Mihr Shahnazaryan2
1Department of Neurosurgery, HELIOS Klinikum Erfurt, Erfurt, Germany; Department of Neurosurgery, Yerevan State Medical University, Yerevan, Armenia.
Objective:
To evaluate efficacy and reliability of intraoperative Doppler sonography in localizing the transverse and sigmoid sinuses during lateral suboccipital craniotomy.
Methods:
A 16-Mhz intraoperative micro-Doppler ultrasound (16Mhz, Multi-Dop pro, Compumedics, Singen, Germany) was applied to detect the medial border of the sigmoid sinus and the inferior border of the transverse sinus in 25 patients. Micro-Doppler measurements were compared with magnetic resonance- and computed tomography-based image guidance (Kolibri, Brainlab, Munich, Germany). Visual detectability of the sinuses with the operating microscope was also documented.
Results:
Inadvertent incision of the transverse or sigmoid sinuses did not occur in any patient when the 2 localizing methods have been used in combination. The mean mismatch of image-guided system and micro-Doppler was 2.64 mm (range, 0-6 mm; standard deviation, 1.55 mm). With the microscope the transverse sinus was invisible in 7 patients, the sigmoid sinus was visually undetectable in 1 case. The micro-Doppler indicated blood flow outside the visible borders of the sinuses in 5 patients.
Conclusions:
A combination of image-guidance and micro-Doppler enhances the accuracy in localizing the margins of the transverse and sigmoid sinuses using the retrosigmoid approach, thus preventing inadvertent injury. The method could potentially be applied during other craniotomies involving the exposure of a venous sinus.
More Related Videos
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called...