Effects of Different Containers on Radioactivity Measurements using a Dose Calibrator with Special Reference to 111In
Yusuke Inoue1, Yutaka Abe2, Kei Kikuchi2
1Department of Diagnostic Radiology, Kitasato University School of Medicine.
Abstract:
Low-energy characteristic x-rays emitted by 111In and 123I sources are easily absorbed by the containers of the sources, affecting radioactivity measurements using a dose calibrator. We examined the effects of different containers on the estimated activities. The radioactivities of 111In, 123I, 201Tl, and 99mTc were measured in containers frequently employed in clinical practice in Japan. The 111In measurements were performed in the vials A and B of the 111In-pentetreotide preparation kit and in the plastic syringe. The activities of 123I-metaiodobenzylguanidine and 201Tl chloride were measured in the prefilled glass syringes and plastic syringes. The milking vial, vial A, vial B, and plastic syringe were used to assay 99mTc. For 111In and 123I, measurements were performed with and without a copper filter. The filter was inserted into the well of the dose calibrator to absorb low-energy x-rays. The relative estimate was defined as the ratio of the activity estimated with the dose calibrator to the standard activity. The estimated activities varied greatly depending on the container when 111In and 123I sources were assayed without the copper filter. The relative estimates of 111In were 0.908, 1.072, and 1.373 in the vial A, vial B, and plastic syringe, respectively. The relative estimates of 123I were 1.052 and 1.352 in the glass syringe and plastic syringe, respectively. Use of the copper filter eliminated the container-dependence in 111In and 123I measurements. Container-dependence was demonstrated in neither 201Tl nor 99mTc measurements. The activities of 111In and 123I estimated with a dose calibrator differ greatly among the containers. Accurate estimation may be attained using the container-specific correction factor or using the copper filter.
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Isotopes and Radioisotopes
An isotope containing...
Radioactive Decay and Radiometric Dating
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...