Reduced metallic artefacts in 3 T knee MRI using fast spin-echo multi-point Dixon compared to fast spin-echo

H J Park1, S Y Lee1, S H Choi1

  • 1Department of Radiology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.

Clinical Radiology
|August 16, 2017
PubMed
Abstract

Insights

The multi-point Dixon MRI technique, specifically its water-only image, significantly reduces metallic artifact size around knee implants compared to fast spin-echo T2-weighted imaging. In-phase images showed no significant difference.

Area of Science:

  • Radiology
  • Medical Imaging
  • Orthopedic Surgery

Background:

  • Metallic implants, such as interference screws used in anterior cruciate ligament (ACL) reconstruction, cause image artifacts in magnetic resonance imaging (MRI).
  • These artifacts can obscure surrounding anatomy, potentially impacting diagnostic accuracy.
  • Fast spin-echo (FSE) T2-weighted imaging (WI) is commonly used but is susceptible to metallic artifacts.

Purpose of the Study:

  • To compare the size of metallic artifacts generated by multi-point Dixon MRI and FSE T2-WI in the knee joint.
  • To evaluate the effectiveness of different multi-point Dixon sequences (water-only and in-phase) in reducing artifact size compared to standard FSE T2-WI.

Main Methods:

  • A retrospective study included 42 patients with ACL reconstruction imaged using 3 Tesla (3T) MRI.
  • Metallic artifact size was measured as the maximal distance of image distortion around interference screws on sagittal images.
  • Artifacts were assessed using FSE T2WI (T2WI, fat-suppressed T2WI) and multi-point Dixon (water-only, in-phase) sequences.

Main Results:

  • The mean artifact distance ranged from 16.6 mm to 24.5 mm across sequences.
  • Multi-point Dixon water-only images showed significantly smaller artifact distances compared to fat-suppressed FSE T2WI (p<0.001).
  • No significant difference in artifact size was found between multi-point Dixon in-phase images and standard FSE T2WI (p>0.05).

Conclusions:

  • The water-only image of the multi-point Dixon technique effectively reduces metallic artifacts compared to fat-suppressed FSE T2WI.
  • The in-phase images of multi-point Dixon do not offer a significant advantage over standard FSE T2WI in artifact reduction.
  • Multi-point Dixon MRI, particularly the water-only sequence, shows promise for improved imaging of metallic implants in the knee.