Cardio-Pulmonary Stethoscope: Clinical Validation With Heart Failure and Hemodialysis Patients
Objective:
The purpose of this study is to evaluate the accuracy of a noninvasive radiofrequency-based device, the Cardio-Pulmonary Stethoscope (CPS), to monitor heart and respiration rates, and detect changes in lung water content in human experiments and clinical trials.
Methods:
Three human populations (healthy subjects ( ), heart failure (), and hemodialysis () patients) were enrolled in this study. The study was conducted at the University of Hawaii and the Queen's Medical Center in Honolulu, HI, USA. Measurement of heart and respiration rates for all patients was compared with standard FDA - approved monitoring methods. For lung water measurements, CPS data were compared with simultaneous pulmonary capillary wedge pressure (PCWP) measurements for heart failure patients, and with change in weight of extracted fluid for hemodialysis patients.
Results:
Statistical correlation methods (Pearson, mixed, and intraclass) were used to compare the data and examine accuracy of CPS results. Results show that heart and respiration rates of all patients have excellent correlation factors, r≥0.9. Comparisons with fluid removed during hemodialysis treatment showed correlation factor of to 1, while PCWP measurements of heart failure patients had correlation factor of to 0.97.
Conclusion:
These results suggest that CPS technology accurately quantifies heart and respiration rates and measure fluid changes in the lungs.
Significance:
The CPS has the potential to accurately monitor lung fluid status noninvasively and continuously in a clinical and outpatient setting. Early and efficient management of lung fluid status is key in managing chronic conditions such heart failure, pulmonary hypertension, and acute respiration distress syndrome.
Related Concept Videos
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Mitral Stenosis II: Clinical features and Diagnostic Tests
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...