Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Comparison of Point-of-Care Versus Laboratory Troponin Testing in an Emergency Department Setting
Michael Juliano1, Courtney Wason2
1U.S. Fleet Forces Command, N01H, 1562 Mitscher Avenue, Suite 250, Norfolk, VA 23551.
Background:
There have been concerns regarding troponin results accuracy between point-of-care (POC) testing preformed in an emergency department (ED) setting and laboratory testing. The purpose of this study was to compare the results of cardiac troponin I testing as concerns POC testing and laboratory analysis as a way to show that these results are interchangeable.
Methods:
A retrospective chart review was performed from October 2012 through September 2013 to identify all patients who presented to the ED that received both a POC and laboratory troponin associated with the same blood draw.
Findings:
A total of 189 patients met inclusion criteria. Three laboratory samples were hemolyzed and not available for comparison. Of the remaining 186 samples, when a positive cutoff value of 0.034 ng/mL was used: 37 POC tests were positive and 149 POC tests were negative, sensitivity 0.85 (95% confidence interval [CI] = 0.69-0.94), specificity 0.98 (95% CI = 0.94-0.99), positive predictive value 91.9% (95% CI = 77-97.9%), and negative predictive value 96% (95% CI = 91.1-98.4%). When a POC troponin positive cutoff value of 0.12 ng/mL was used: 20 POC tests were positive and 164 were negative: sensitivity 1.0 (95% CI = 0.8-1.0), specificity 0.99 (95% CI = 0.95-1.0), positive predictive value 91% (95% CI = 69-98.4%), and negative predictive value 100% (95% CI = 97.1-100%).
Discussion:
In an ED setting; POC testing is accurate and correlates well with laboratory testing. Considering that the POC analysis takes a fraction of time to yield results, the validity of its data is important. The results of this study show that the POC system is a suitable test for rapid evaluation of patients presenting to the ED. This correlation increased with a higher testing positive cutoff value (0.12 versus 0.034 ng/mL). These data can assist the emergency physician with rapid identification of evidence of cardiac injury and be assured that the results are accurate.
Recommendations:
POC lab testing should be considered whenever possible to provide the fastest troponin results.
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...