Human Factor and Usability Testing of a Binocular Optical Coherence Tomography System
Reena Chopra1, Pádraig J Mulholland1,2, Adam M Dubis1
1NIHR Biomedical Research Centre at Moorfields Eye Hospital NHS Foundation Trust and UCL Institute of Ophthalmology, London, UK.
This study shows a new binocular optical coherence tomography (OCT) device is easy to use for elderly and visually impaired patients. The automated eye exam system has potential to improve eye care efficiency and quality.
Area of Science:
- Ophthalmology
- Medical Technology
- Usability Engineering
Background:
- Automated eye examination systems aim to improve efficiency and accessibility in clinical settings.
- Optical coherence tomography (OCT) is a key imaging modality in ophthalmology.
- Usability testing is crucial for the successful clinical integration of new medical devices, especially for diverse patient populations.
Purpose of the Study:
- To evaluate the usability of a novel binocular optical coherence tomography (OCT) prototype.
- To assess the potential for automated eye examinations in a clinical environment.
- To identify user errors and subjective feedback, particularly from elderly and visually impaired individuals.
Main Methods:
- Usability testing was conducted with 45 participants with chronic eye disease and 15 healthy controls.
- The prototype performed automated 'whole-eye' OCT, ocular motility, visual acuity, perimetry, and pupillometry.
- Post-examination interviews gathered subjective feedback on ease of use and appeal.
Main Results:
- All participants successfully completed the automated eye examination without assistance.
- A high percentage of users (81% with eye disease, 79% healthy) found the prototype easier to use than smartphones.
- 86% of participants considered the device appealing for clinical use, with no significant time difference between groups.
Conclusions:
- The developed binocular OCT system is usable and can be administered in an automated fashion.
- This technology shows potential to enhance the efficiency and quality of comprehensive eye examinations.
- Further development is guided by identified areas for improvement to facilitate clinical translation.