Falsely high B-type natriuretic peptide concentration in patients without heart failure attributed to AxSYM assay:
Cai De Jin1,2, Moo Hyun Kim1,2, Long Zhe Guo1,2
1Department of Cardiology, College of Medicine, Dong-A University, Busan, Korea.
Abstract:
We report a case series of eight subjects complaining of non-specific chest pain without heart failure, but with apparent high concentrations of plasma B-type natriuretic peptide (BNP). No positive clinical characteristics were identified in physical examinations, cardiac imaging, laboratory findings, or pulmonary function tests. However, we observed unusually high BNP values when analysing blood samples of the patients using the AxSYM assay, and this was not supported by readings from Triage® or ADVIA Centaur® assays on the same samples, which showed BNP within the normal range. We believe that the possibility for false readings for high BNP levels in healthy individuals measured by AxSYM assay should be taken into account by physicians in clinical practice to avoid medical errors.
Related Concept Videos
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Heart Failure II: Pathophysiology
Errors occurring during blood pressure monitoring
Several factors...
Heart Failure Drugs: Inhibitors of Renin-Angiotensin System
Nephrotic Syndrome II : Assessment and Medical Management
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...