Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
Toward image quality assessment in mammography using model observers: Detection of a calcification-like object
Ramona W Bouwman1, Alistair Mackenzie2, Ruben E van Engen1
1Dutch Expert Centre for Screening (LRCB), Radboud University Medical Center, PO Box 6873, 6503 GJ, Nijmegen, The Netherlands.
Purpose:
Model observers (MOs) are of interest in the field of medical imaging to assess image quality. However, before procedures using MOs can be proposed in quality control guidelines for mammography systems, we need to know whether MOs are sensitive to changes in image quality and correlations in background structure. Therefore, as a proof of principle, in this study human and model observer (MO) performance are compared for the detection of calcification-like objects using different background structures and image quality levels of unprocessed mammography images.
Method:
Three different phantoms, homogeneous polymethyl methacrylate, BR3D slabs with swirled patterns (CIRS, Norfolk, VA, USA), and a prototype anthropomorphic breast phantom (Institute of Medical Physics and Radiation Protection, Technische Hochschule Mittelhessen, Germany) were imaged on an Amulet Innovality (FujiFilm, Tokyo, Japan) mammographic X-ray unit. Because the complexities of the structures of these three phantoms were different and not optimized to match the characteristics of real mammographic images, image processing was not applied in this study. In addition, real mammograms were acquired on the same system. Regions of interest (ROIs) were extracted from each image. In half of the ROIs, a 0.25-mm diameter disk was inserted at four different contrast levels to represent a calcification-like object. Each ROI was then modified, so four image qualities relevant for mammography were simulated. The signal-present and signal-absent ROIs were evaluated by a non-pre-whitening model observer with eye filter (NPWE) and a channelized Hotelling observer (CHO) using dense difference of Gaussian channels. The ROIs were also evaluated by human observers in a two alternative forced choice experiment. Detectability results for the human and model observer experiments were correlated using a mixed-effect regression model. Threshold disk contrasts for human and predicted human observer performance based on the NPWE MO and CHO were estimated.
Results:
Global trends in threshold contrast were similar for the different background structures, but absolute contrast threshold levels differed. Contrast thresholds tended to be lower in ROIs from simple phantoms compared with ROIs from real mammographic images. The correlation between human and model observer performance was not affected by the range of image quality levels studied.
Conclusions:
The correlation between human and model observer performance does not depend on image quality. This is a promising outcome for the use of model observers in image quality analysis and allows for subsequent research toward the development of MO-based quality control procedures and guidelines.
Insights
Model observers (MOs) show consistent performance correlation with human observers across varying mammography image quality levels. This finding supports the development of MO-based quality control procedures for mammography systems.
Area of Science:
- Medical Imaging
- Radiology
- Image Quality Assessment
Background:
- Model observers (MOs) are crucial for assessing image quality in medical imaging.
- Their sensitivity to image quality variations and background structures needs validation before integration into mammography quality control.
Purpose of the Study:
- To compare human and model observer performance in detecting calcification-like objects.
- To evaluate sensitivity to background structures and image quality variations in mammography.
Main Methods:
- Utilized homogeneous, BR3D, and anthropomorphic phantoms, alongside real mammograms.
- Simulated four image quality levels and introduced calcification-like objects.
- Evaluated performance using non-pre-whitening (NPWE) and channelized Hotelling observers (CHO), alongside human observers in a two-alternative forced choice experiment.
Main Results:
- Global trends in threshold contrast were similar across background structures, but absolute levels varied.
- Contrast thresholds were lower in simpler phantom ROIs compared to real mammograms.
- The correlation between human and model observer performance remained unaffected by image quality variations.
Conclusions:
- Human and model observer performance correlation is independent of image quality.
- This supports the use of MOs for image quality analysis.
- Enables future research into MO-based quality control procedures for mammography.