Head and neck angiography at 70 kVp with a third-generation dual-source CT system in patients: comparison with
Yu Chen1, Xiaobo Zhang1, Huadan Xue1
1Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No.1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China.
Purpose:
This study was conducted in order to evaluate the image quality of 70 kVp and 25 mL contrast medium (CM) volume for head and neck computed tomographic angiography (CTA) and assess the diagnostic accuracy for arterial stenosis.
Methods:
Fifty patients were prospectively divided into two groups randomly: group A (n = 25), 70 kVp with 25 mL CM, and group B (n = 25), 100 kVp with 40 mL CM. CT attenuation values, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the shoulder, neck, and cerebral arteries were measured for objective image quality. Subjective image quality of the shoulder and cerebral arteries was also evaluated. For patients undergoing digital subtracted angiography (DSA), diagnostic accuracy of CTA was assessed with DSA as reference standard.
Results:
The SNRs of the shoulder, neck, and cerebral arteries in group A were higher than those in group B (P < 0.05). The CNRs of the shoulder and neck arteries in group A were higher than those in group B (P < 0.05). There was no significant difference in subjective image quality of arteries between group A and group B (P > 0.05). The accuracy was noted as 94.0% (156/166) in group A and 97.1% (134/138) in group B for ≥ 50% stenosis. The accuracy of intracranial arterial stenosis was lower than that of extracranial arterial stenosis in group A. The radiation dose of group A was significantly decreased by 56% than that of group B.
Conclusion:
Head and neck CTA at 70 kVp using 25 mL CM can obtain diagnostic image quality with lower radiation dose while maintaining high accuracy in detecting the arterial stenosis compared with the 100-kVp and 40-mL CM.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...