The AB Portable Driver Generates Higher Drive-Line Pressures Possibly Leading to Accelerated Hemolysis
Motohiko Goda1, Naoto Yabu1, Norihisa Tominaga1
1Department of Cardiovascular Surgery, Yokohama City University, Yokohama, Japan.
Abstract:
The AB5000 Circulatory Support System is paracorporeal pulsatile ventricular assist device. The AB Portable Driver is a portable console for this system. We experienced two cases with accelerated hemolysis while receiving support by the AB Portable Driver. The purpose of this study was to clarify the mechanical differences associated with the hemolysis between the AB5000 console and the AB Portable Driver. The mock circulatory system modeled by an AB5000 ventricle and a blood sampling bag of vinyl chloride was run with an AB5000 console or AB Portable Driver. The peak drive-line pressure, the mean arterial cannula pressure and the pumping rate of the VAD were recorded. The AB5000 console generated a peak drive-line pressure of 280-300 mm Hg in LVAD mode and 210-220 mm Hg in RVAD mode, approximately 100 mm Hg lower than officially documented. In contrast, the AB Portable Driver generated pressures of 310-330 mm Hg in LVAD mode and 230-250 mm Hg in RVAD mode, 65-95 mm Hg higher than officially documented. The AB Portable Driver console generates higher drive-line pressures than the AB5000 console, possibly explaining the accelerated hemolysis.
More Related Videos
11:12Ex Vivo Red Blood Cell Hemolysis Assay for the Evaluation of pH-responsive Endosomolytic Agents for Cytosolic Delivery of Biomacromolecular Drugs
Published on: March 9, 2013
09:59Laser-free Hydroxyl Radical Protein Footprinting to Perform Higher Order Structural Analysis of Proteins
Published on: June 4, 2021