Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Tonometry after Intrastromal Corneal Ring Segments for Keratoconus
Paula Arribas-Pardo1, Carmen Mendez-Hernandez, Ricardo Cuiña-Sardiña
11Ophthalmology Department, Hospital Clinico San Carlos, Institute of Health Care Research of the Hospital Clinico San Carlos (IdISSC), University Complutense of Madrid, Madrid, Spain 2Cooperative Research Network in Ophthalmology (RETICS), Instituto de Salud Carlos III, Madrid, Spain *paula_arribas_pardo@hotmail.com.
Accurate intraocular pressure (IOP) measurement after intrastromal corneal ring segments (ICRS) implantation is crucial. The iCare Pro and TonoPen XL tonometers demonstrated the best agreement with Goldmann applanation tonometry (GAT) in ICRS patients.
Area of Science:
- Ophthalmology
- Corneal Surgery
- Ophthalmic Diagnostics
Background:
- Intrastromal corneal ring segments (ICRS) alter corneal morphology, challenging reliable intraocular pressure (IOP) measurement.
- Accurate IOP assessment is vital for managing conditions like corneal ectasia post-ICRS implantation.
Purpose of the Study:
- To compare IOP measurements from different tonometers against Goldmann applanation tonometry (GAT) in patients with ICRS.
- To evaluate the influence of central corneal thickness (CCT), corneal curvature, and astigmatism on IOP readings after ICRS.
Main Methods:
- Cross-sectional study involving 91 eyes of 91 patients previously implanted with ICRS.
- IOP measured using TonoPen XL, Pascal dynamic contour tonometer (DCT), and iCare Pro, with GAT as the reference.
- Statistical analysis included Bland-Altman method, intraclass correlation coefficients, and multiple linear regression.
Main Results:
- Mean IOP differences (GAT vs. tonometer): TonoPen XL (-0.8 ± 3.07 mmHg), DCT (-1.0 ± 3.26 mmHg), iCare Pro (0.8 ± 2.92 mmHg).
- Central corneal thickness (CCT) was identified as a confounding factor for all tonometers except DCT.
- DCT-IOP measurements were not affected by corneal topographic factors.
Conclusions:
- The iCare Pro and TonoPen XL showed the highest agreement with GAT for IOP measurement in ICRS patients.
- The dynamic contour tonometer (DCT) provided IOP readings unaffected by corneal parameters but tended to overestimate GAT values.
More Related Videos
07:51Full-Field Optical Coherence Microscopy for Histology-Like Analysis of Stromal Features in Corneal Grafts
Published on: October 21, 2022
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment