Validation of a New System Using Tracheal Body Sound and Movement Data for Automated Apnea-Hypopnea Index Estimation
Christoph Kalkbrenner1, Manuel Eichenlaub2, Stefan Rüdiger3
1Faculty of Medical Engineering, University of Applied Science Ulm, Ulm, Germany.
Study Objectives:
The current gold standard for assessment of obstructive sleep apnea is the in-laboratory polysomnography. This approach has high costs and inconveniences the patient, whereas alternative ambulatory systems are limited by reduced diagnostic abilities (type 4 monitors, 1 or 2 channels) or extensive setup (type 3 monitors, at least 4 channels). The current study therefore aims to validate a simplified automated type 4 monitoring system using tracheal body sound and movement data.
Methods:
Data from 60 subjects were recorded at the University Hospital Ulm. All subjects have been regular patients referred to the sleep center with suspicion of sleep-related breathing disorders. Four recordings were excluded because of faulty data. The study was of prospective design. Subjects underwent a full-night screening using diagnostic in-laboratory polysomnography and the new monitoring system concurrently. The apnea-hypopnea index (AHI) was scored blindly by a medical technician using in-laboratory polysomnography (AHIPSG). A unique algorithm was developed to estimate the apneahypopnea index (AHIest) using the new sleep monitor.
Results:
AHIest strongly correlates with AHIPSG (r2 = .9871). A mean ± 1.96 standard deviation difference between AHIest and AHIPSG of 1.2 ± 5.14 was achieved. In terms of classifying subjects into groups of mild, moderate, and severe sleep apnea, the evaluated new sleep monitor shows a strong correlation with the results obtained by polysomnography (Cohen kappa > 0.81). These results outperform previously introduced similar approaches.
Conclusions:
The proposed sleep monitor accurately estimates AHI and diagnoses sleep apnea and its severity. This minimalistic approach may address the need for a simple yet reliable diagnosis of sleep apnea in an ambulatory setting.
Clinical Trial Registration:
Trial name: Validation of a new method for ambulant diagnosis of sleep related breathing disorders using body sound; URL: https://drks-neu.uniklinik-freiburg.de/drks_web/navigate.do?navigationId=trial.HTML&TRIAL_ID=DRKS00011195; Identifier: DRKS00011195.
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Sleep Apnea
The condition is more prevalent among...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.