Diagnosis of prosthetic joint infection with alpha-defensin using a lateral flow device: a multicentre study

P Berger1, M Van Cauter2, R Driesen3

  • 1University Hospitals Leuven, Weligerveld 1, 3212 Pellenberg, Belgium.

The Bone & Joint Journal
|September 2, 2017
PubMed
Abstract

Insights

The Synovasure lateral flow device accurately detects alpha-defensin proteins in synovial fluid, offering a reliable method for diagnosing prosthetic joint infections (PJI). This test shows excellent performance for painful knee and hip arthroplasties.

Area of Science:

  • Orthopaedic surgery
  • Infectious disease diagnostics
  • Biomarker analysis

Background:

  • Prosthetic joint infection (PJI) is a significant complication following joint replacement surgery.
  • Accurate and timely diagnosis of PJI is crucial for effective patient management and treatment.
  • Current diagnostic criteria, such as those from the Musculoskeletal Infection Society (MSIS), are essential but can be complex.

Purpose of the Study:

  • To evaluate the diagnostic accuracy of the Synovasure lateral flow device for detecting alpha-defensin proteins in synovial fluid.
  • To compare the Synovasure test's performance against established MSIS criteria for diagnosing PJI.
  • To assess the reliability of alpha-defensin testing in patients with painful prosthetic joints.

Main Methods:

  • A prospective multicentre study involving 121 patients with painful knee or hip arthroplasties.
  • Synovial fluid samples were analyzed using the Synovasure lateral flow device.
  • Results were compared with diagnoses based on the MSIS criteria for PJI.

Main Results:

  • The Synovasure test demonstrated high sensitivity (97.1%) and specificity (96.6%) for PJI detection.
  • Positive and negative predictive values were 91.7% and 98.8%, respectively.
  • Receiver operator characteristics analysis showed an area under the curve of 0.97, indicating excellent diagnostic performance.

Conclusions:

  • The Synovasure test exhibits excellent diagnostic performance for confirming or excluding PJI.
  • This alpha-defensin test is a promising tool for managing painful knee and hip arthroplasties.
  • The Synovasure test should be considered a valuable addition to the diagnostic armamentarium for PJIs.