Flow characteristics of urethral catheters of the same caliber vary between manufacturers
Carrie A Stewart1, Eiichiro Yamaguchi2, Jessica Teixeira Vaz3
1Tulane University School of Medicine, New Orleans, LA, USA.
Background:
Clean intermittent catheterization (CIC) is frequently prescribed for bladder dysfunction, either per urethra or via a continent catheterizable channel. Small catheters may be required for infants or continent channels. Success with CIC is highly dependent upon patient and family compliance. The urinary flow rate through the catheter is an important factor, which can decrease CIC time and improve quality of life. There is little objective information regarding flow rate through urinary catheters to guide catheter recommendation or prescription. Clinically, we noted that there was a difference in flow among catheter brands, and we questioned if catheters of the same-labeled diameter exhibit the same flow characteristics, which could have implications for catheter selection.
Methods:
Twenty-one commercially available male pediatric urinary catheters from nine brands were tested (11 straight tip, 10 coude tip). Nine of the 21 tested catheters had a hydrophilic coating. All tested catheters shared a 10F outer diameter. For each, microscopic imaging and a precision caliper were used to measure the inner diameter and tip inlet area. A hydraulic system modified from ASTM standard testing specifications was used to simulate bladder catheterization. Measurement of each catheter was repeated five times using three different static hydraulic pressures (20, 40 and 50 cmH2O). Catheter flow rate and structural measurements were identified and the fastest and slowest of the catheters are presented in the table. The variable flow rates between brands were due to the differences in catheter structural characteristics such as the inner diameter (ID) and the tip inlet area to inner lumen area ratio (AR). The maximum variation of flow rate of all tested 10F catheters was 48%, ID varied up to 22%, from 1.71 to 2.11 mm or 5.13-6.33F. AR varied up to 166%. The table delineates the fastest and slowest rates at three measured pressures. The outer diameter labeled 10F on packaging was true to size.
Conclusions:
Based on packaging information, providers, and patients are unable to predict urinary flow through a catheter and thus use information regarding flow rate to guide catheter selection. This information cannot be calculated based on ideal flow calculations and could be listed on packaging to assist physicians and families in selecting the optimal urinary catheter for CIC.
More Related Videos
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Urethra
The anatomy of the urethra differs between males and females. In females, the urethra is short, measuring about 3–4 cm in length, and opens anterior to the vaginal opening. In males, the urethra is longer and passes through the penis, serving dual purposes: expelling urine and ejaculating semen. The male...
Ureters
Pipe Flowrate Measurement
The orifice meter is a simple,...
Anatomy of the Genitourinary System II: Bladder and Urethra
General Characteristics of Pipe Flow I
The classification of fluid...