Related Experiment Video
Updated: Feb 23, 2026

Modeling and Experimental Analysis of the Single-Shaft Coaxial Motor-Pump Assembly in Electrohydrostatic Actuators
Published on: June 13, 2022
Clinical Experience with the Nikkiso Centrifugal Pump
Yu Shomura1, Takatsugu Shimono1, Koji Onoda1
1Department of Thoracic and Cardiovascular Surgery, Mie University School of Medicine, Mie, Japan.
The Nikkiso HPM-15 centrifugal pump demonstrates safety and ease of use in cardiopulmonary bypass (CPB) and percutaneous cardiopulmonary support (PCPS). Clinical results show stable hemodynamics and minimal impact on organ function, making it a useful device for these procedures.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Medical Devices
Background:
- The Nikkiso HPM-15 is a compact centrifugal pump previously evaluated for clinical use.
- Advancements include a newly developed controller for enhanced functionality.
Purpose of the Study:
- To report clinical experiences with the Nikkiso HPM-15 centrifugal pump in cardiopulmonary bypass (CPB) and percutaneous cardiopulmonary support (PCPS).
- To assess the pump's safety, ease of operation, and hemodynamic performance.
Main Methods:
- Clinical evaluation of the Nikkiso HPM-15 pump in 23 patients undergoing CPB (22 patients) or PCPS (1 patient).
- Assessment of circuit preparation, deaeration, hemodynamic stability, serum-free hemoglobin, platelet count, urine output, and renal/liver function.
Main Results:
- The pump facilitated easy circuit preparation and effective deaeration.
- Stable hemodynamics were maintained throughout CPB and PCPS procedures.
- Minimal increase in serum-free hemoglobin, temporary decrease in platelet count (recovering within 7 days), and high urine output were observed.
- No renal or liver function abnormalities occurred.
Conclusions:
- The Nikkiso HPM-15 centrifugal pump is safe and user-friendly.
- The device is effective and suitable for clinical application in CPB and PCPS procedures.
Related Concept Videos
Centrifugal Force
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
Centrifugation
Application of the Energy Equation
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...
High-Performance Liquid Chromatography: Elution Process