Study of Secondary Flow in Centrifugal Blood Pumps Using a Flow Visualization Method with a High-Speed Video Camera
Ichiro Sakuma1, Yasuhiro Fukui1, Takeyoshi Dohi1
1Department of Applied Electronic Engineering, Faculty of Science and Engineering, Tokyo Denki University SaitamaDepartment of Precision Machinery Engineering, Faculty of Engineering, The University of Tokyo, Tokyo, Japan.
Abstract:
Four pump models with different vane configurations were evaluated with flow visualization techniques using a high-speed video camera. These models also were evaluated through in vivo hemolysis tests using bovine blood. The impeller having the greatest fluid velocity relative to the impeller, the largest velocity variance, and the most irregular local flow patterns in the flow passage caused the most hemolysis. Even if the pumps were operated at almost the same speed (rpm) at the same output, the impeller showing more irregular flow patterns had a statistically greater rate of hemolysis. This fact confirms that the existence of local irregular flow patterns in a centrifugal blood pump deteriorates its hemolytic performance. Thus, to optimize the design of the pump, it is very important to examine the secondary flow patterns in the centrifugal blood pump in detail using flow visualization with a high-speed video camera.
Related Concept Videos
Blood Flow
Rapidly Varying Flow
Laminar and Turbulent Flow
Steady, Laminar Flow in Circular Tubes
Bernoulli's Equation for Flow Normal to a Streamline
The pressure difference depends on the fluid's velocity and radius of curvature. The pressure variation is minimal in flows with nearly straight streamlines. However, the...
Steady, Laminar Flow Between Parallel Plates