Measurement of Milwaukee Brace Pad Pressure in Adolescent Round Back Deformity Treatment
Related Experiment Video
Updated: Feb 23, 2026

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Taher Babaee1, Mojtaba Kamyab1, Amir Ahmadi2
1Department of Orthotics and Prosthetics, School of Rehabilitation Sciences, Iran University of Medical sciences, Tehran, Iran.
Study Design:
In this prospective study, we measured the pad pressures of the Milwaukee brace in adolescent hyperkyphosis treatment.
Purpose:
We evaluated the skin-brace interface forces exerted by the main pads of the Milwaukee brace.
Overview Of Literature:
A fundamental factor associated with brace effectiveness in spinal deformity is pad force adjustment. However, few studies have evaluated the in-brace force magnitude and its effect on curve correction.
Methods:
Interface forces at four pads of the Milwaukee brace were measured in 73 patients withround back deformity (mean age, 14.04±1.97 years [range, 10-18]; mean initial Cobb angle,67.70°±9.23° [range, 50°-86°]). We used a modified aneroid sphygmomanometer to measure the shoulder and kyphosis pad pressures. Each patient underwent measurement in the standing and sitting positions during inhalation/exhalation.
Results:
The mean pad pressures were significantly higher in the standing than in thesitting position, and significantly higher pressures were observed during inhalation compared toexhalation (p=0.001).There were no statistically significant differences between right and left shoulder pad pressures (p>0.05); however, the pressure differences between the right and left kyphosis pads were statistically significant (p<0.05). In a comparison of corrective forces with bracing for less or more than 6 months, corrective force was larger with bracing for less than 6 months (p=0.02). In the standing position, there were no statistically significant correlations between pad pressures and kyphosis curve correction.
Conclusions:
In the sitting position, there was a trend toward lower forces at the skin-brace interface; therefore, brace adjustment in the standing position may be useful and more effective. There was no significant correlation between the magnitude of the pad pressures and the degree of in-brace curve correction.