Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
Development and validation of an impedance biosensor for point-of-care detection of vascular cell adhesion molecule-1
Anjan Panneer Selvam1, Andi Wangzhou1,2, Michael Jacobs1
1Department of Bioengineering, University of Texas at Dallas, Richardson, TX 75080, USA.
Aim:
Systemic lupus erythematosus is an autoimmune disease that requires chronic monitoring. In this study, we demonstrate a proof-of-concept study of a highly attractive noninvasive strategy for monitoring systemic lupus erythematosus through biomarker quantification.
Results:
This sensor technology requires 50 μl of urine to detect and quantify vascular cell adhesion molecule-1 in 15 min. The sensor used nonfaradaic detection to demonstrate performance with and without detection antibody. Binding of immunoassay and target biomarkers were quantified with an impedance electrical immunoassay and correlated with an equivalent circuit.
Conclusion:
The novel sensor technology demonstrates detection in the range of 8 fg/ml to 800 pg/ml and comparative analysis with ELISA platforms was performed for 12 patient urine samples.