Method and Instrumented Fixture for Femoral Fracture Testing in a Sideways Fall-on-the-Hip Position
Published on: August 17, 2017
Comparison of 2 plating techniques for lateral clavicle fractures, using a new standardized biomechanical testing
Cyrill Suter1, Martin Majewski1, Andrej M Nowakowski1
1Orthopaedic Department, University of Basel, Basel - Switzerland.
Background:
Conservative treatment of lateral clavicle fractures (Neer type II) often ends with unsatisfactory results for patients. Operative treatment is thus gaining acceptance. A retrospective study found success using a double plate technique for these injuries. For the current study, a standardized testing setup was developed to compare the mechanical properties of single versus double plate technique for lateral clavicle fractures.
Methods:
Six synthetic bones were tested for each technique. Neer type IIB fractures were created using computer-aided design (CAD). Fatigue testing was carried out with a cantilever bending test. Parameters measured were cycles undergone, failure load and stiffness at the point of failure.
Results:
The standardized testing setup was able to provide reproducible failures. The double plate technique lasted about 16,000 more cycles and withstood an additional 22.4 N on average.
Conclusions:
The new modified standardized testing setup produced reproducible fixation failures for both clavicle fracture fixation techniques. The double plate technique seems to be mechanically superior to the single plate technique.