Reproducibility and normal values of static pupil diameters
Simon Schröder1, Ekaterina Chashchina1, Edgar Janunts1
11 Department of Experimental Ophthalmology, Saarland University, Homburg - Germany.
European Journal of Ophthalmology
|September 9, 2017
Summary
This study establishes normal pupil diameter values using the PupilX infrared pupillometer across different light conditions. The device demonstrated high reproducibility for measuring pupil size and anisocoria in normal subjects.
Area of Science:
- Ophthalmology
- Optometry
- Biomedical Engineering
Background:
- Pupil diameter is a critical biometric parameter influenced by age and illumination.
- Accurate measurement of pupil size is essential for various clinical and research applications.
- Infrared pupillometry offers a non-invasive method for assessing pupillary responses.
Purpose of the Study:
- To determine normative static pupil diameter values using the PupilX infrared pupillometer.
- To evaluate the reproducibility of pupil diameter measurements obtained with the PupilX device.
- To investigate age and light-related variations in pupil size.
Main Methods:
- 91 participants with normal eyes underwent binocular infrared pupillometry using PupilX.
- Pupil diameters were measured under scotopic (0 lx), mesopic (1 lx), and photopic (16 lx) illumination.
- Each measurement was repeated five times to assess device repeatability.
Main Results:
- Mean pupil diameters were 6.5 mm (scotopic), 5.5 mm (mesopic), and 4.03 mm (photopic).
- No significant difference in pupil diameter was observed between left and right eyes.
- The PupilX device exhibited high repeatability, with unsigned anisocoria consistently below 0.3 mm and overall repeatability better than 0.2 mm.
Conclusions:
- The study provides valuable reference data for normal pupil diameters measured with the PupilX digital pupillometer.
- These findings contribute to understanding age and light-dependent changes in pupil size.
- The PupilX device is a reliable tool for reproducible pupil diameter measurements.