Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
An in vitro simulation method for the tribological assessment of complete natural hip joints
Dawn Groves1, John Fisher1, Sophie Williams1
1Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom.
Abstract:
The use of hip joint simulators to evaluate the tribological performance of total hip replacements is widely reported in the literature, however, in vitro simulation studies investigating the tribology of the natural hip joint are limited with heterogeneous methodologies reported. An in vitro simulation system for the complete natural hip joint, enabling the acetabulum and femoral head to be positioned with different orientations whilst maintaining the correct joint centre of rotation, was successfully developed for this study. The efficacy of the simulation system was assessed by testing complete, matched natural porcine hip joints and porcine hip hemiarthroplasty joints in a pendulum friction simulator. The results showed evidence of biphasic lubrication, with a non-linear increase in friction being observed in both groups. Lower overall mean friction factor values in the complete natural joint group that increased at a lower rate over time, suggest that the exudation of fluid and transition to solid phase lubrication occurred more slowly in the complete natural hip joint compared to the hip hemiarthroplasty joint. It is envisaged that this methodology will be used to investigate morphological risk factors for developing hip osteoarthritis, as well as the effectiveness of early interventional treatments for degenerative hip disease.