Related Experiment Video
Updated: Feb 23, 2026

Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular
Michael Lasta1,2, Kata Miháltz1,2, Illés Kovács3
1Department of Ophthalmology, Hospital Hietzing, Vienna, Austria.
Purpose:
To compare the effect of spherical aberration on optical quality in eyes with two different aspherical intraocular lenses.
Methods:
120 eyes of 60 patients underwent phacoemulsification. In patients' eyes, an aberration-free IOL (Aspira-aA; Human Optics) or an aberration-correcting aspherical IOL (Tecnis ZCB00; Abott Medical Optics) was randomly implanted. After surgery, contrast sensitivity and wavefront measurements as well as tilt and decentration measurements were performed.
Results:
Contrast sensitivity was significantly higher in eyes with Aspira lens under mesopic conditions with 12 cycles per degree (CPD) and under photopic conditions with 18 CPD (p = 0.02). Wavefront measurements showed a higher total spherical aberration with a minimal pupil size of 4 mm in the Aspira group (0.05 ± 0.03) than in the Tecnis group (0.03 ± 0.02) (p = 0.001). Strehl ratio was higher in eyes with Tecnis (0.28 ± 0.17) with a minimal pupil size larger than 5 mm than that with Aspira (0.16 ± 0.14) (p = 0.04). In pupils with a minimum diameter of 4 mm spherical aberration had a significant effect on Strehl ratio, but not in pupils with a diameter less than 4 mm.
Conclusions:
Optical quality was better in eyes with the aberration-correcting Tecnis IOL when pupils were large. In contrast, this could not be shown in eyes with pupils under 4 mm or larger. This trial is registered with Clinicaltrials.gov NCT03224728.
More Related Videos
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...