Objective comparison of high-contrast spatial resolution and low-contrast detectability for various clinical
Damien Racine1, Anaïs Viry1, Fabio Becce2
1Institute of Radiation Physics, Lausanne University Hospital, Rue du Grand-Pré 1, 1007, Lausanne, Switzerland.
Purpose:
We sought to compare objectively computed tomography (CT) scanner performance for three clinically relevant protocols using a task-based image quality assessment method in order to assess the potential for radiation dose reduction.
Methods:
Four CT scanners released between 2003 and 2007 by different manufacturers were compared with four CT scanners released between 2012 and 2014 by the same manufacturers using ideal linear model observers (MO): prewhitening (PW) MO and channelized Hotelling (CHO) MO with Laguerre-Gauss channels for high-contrast spatial resolution and low-contrast detectability (LCD) performance, respectively. High-contrast spatial resolution was assessed using a custom-made phantom that enabled the computation of the target transfer function (TTF) and noise power spectrum (NPS). Low-contrast detectability was assessed using a commercially available anthropomorphic abdominal phantom providing equivalent diameters of 24, 29.6, and 34.6 cm. Three protocols were reviewed: a head (trauma) and an abdominal (urinary stones) protocol were applied to assess high-contrast spatial resolution performance; and another abdominal (focal liver lesions) protocol was applied for LCD. The liver protocol was tested using fixed and modulated tube currents. The PW MO was proposed for assessing high-contrast detectability performance of the various CT scanners.
Results:
Compared with older generation CT scanners, three newer systems displayed significant improvements in high-contrast detectability over that of their predecessors. A fourth, newer system had lower performance. The CHO MO was appropriate for assessing LCD performance and revealed that an excellent level of image quality could be obtained with newer scanners at significantly lower dose levels.
Conclusions:
This study shows that MO can objectively benchmark CT scanners using a task-based image quality method, thus helping to estimate the potential for further dose reductions offered by the latest systems. Such an approach may be useful for adequately and quantitatively comparing clinically relevant image quality among various scanners.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT