Low-dose 90Y PET/CT imaging optimized for lesion detectability and quantitative accuracy: a phantom study to assess the feasibility of pretherapy imaging to plan the therapeutic dose
Related Experiment Video
Updated: Feb 23, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Maryam Khazaee1, Alireza Kamali-Asl, Parham Geramifar
1aDepartment of Medical Radiation Engineering, Shahid Beheshti University bResearch Center for Nuclear Medicine, Shariati Hospital, Tehran University of Medical Sciences, Tehran, Iran cDepartment of Radiology dDepartment of Electrical and Computer Engineering, Johns Hopkins University, Baltimore, Maryland, USA.
Objective:
The overall aim of this work is to optimize the reconstruction parameters for low-dose yttrium-90 (Y) PET/CT imaging, and to determine Y minimum detectable activity, in an endeavor to investigate the feasibility of performing low-dose Y imaging in-vivo to plan the therapeutic dose in radioembolization.
Materials And Methods:
This study was carried out using a Siemens Biograph 6 True Point PET/CT scanner. A Jaszczak phantom containing five hot syringes was imaged serially over 15 days. For 128 reconstruction parameters/algorithms, detectability performance and quantitative accuracy were evaluated using the contrast-to-noise ratio and the recovery coefficient, respectively.
Results:
For activity concentrations greater than 2.5 MBq/ml, the linearity of the scanner was confirmed while the corresponding relative error was below 10%. Reconstructions with smaller numbers of iterations and smoother filters led to higher detectability performance, irrespective of the activity concentration and lesion size. In this study, the minimum detectable activity was found to be 3.28±10% MBq/ml using the optimized reconstruction parameters. Although the recovered activities were generally underestimated, for lesions with activity concentration greater than 4 MBq/ml, the amount of underestimation is limited to -15% for optimized reconstructions.
Conclusion:
Y PET/CT imaging, even with a low activity concentration, is feasible for depicting the distribution of Y implanted microspheres using optimized reconstruction parameters. As such, in-vivo PET/CT imaging of low-dose Y in the pretherapeutic stage may be feasible and fruitful to optimally plan the therapeutic activity delivered to patients undergoing radioembolization.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET