Optically controlled on-coil amplifier with RF monitoring feedback
Natalia Gudino1, Jacco A de Zwart1, Qi Duan1
1Laboratory of Functional and Molecular Imaging, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
Purpose:
To develop a new optically controlled on-coil amplifier that facilitates safe use of multi-channel radiofrequency (RF) transmission in MRI by real-time monitoring of signal phase and amplitude.
Methods:
Monitoring was carried out with a 4-channel prototype system by sensing, down sampling, digitizing, and optically transmitting the RF transmit signal to a remote PC to control the amplifiers. Performance was evaluated with benchtop and 7 T MRI experiments.
Results:
Monitored amplitude and phase were stable across repetitions and had standard deviations of 0.061 microT and 0.0073 rad, respectively. The feedback system allowed inter-channel phase and B1 amplitude to be adjusted within two iterations. MRI experiments demonstrated the feasibility of this approach to perform safe and accurate multi-channel RF transmission and monitoring at high field.
Conclusion:
We demonstrated a 4-channel transceiver system based on optically controlled on-coil amplifiers with RF signal monitoring and feedback control. The approach allows the safe and precise control of RF transmission fields, required to achieve uniform excitation at high field. Magn Reson Med 79:2833-2841, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Related Concept Videos
Cascaded Op Amps
In a cascaded system, each op-amp is referred to as a stage. The output of one stage drives the input of the subsequent stage. As the input signal passes through...
MOSFET Amplifiers
Small-Signal Analysis of MOSFET Amplifiers
Operational Amplifiers
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Second-order Op Amp Circuits
The analysis of such circuits follows a systematic approach, similar to the second-order RLC circuits. In practical scenarios, bulky inductors are rarely employed due to their size and weight. This means...