[Preclinical tests of additive materials for personified endoprostheses of hand joints]
Related Experiment Video
Updated: Feb 22, 2026

Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
G P Kotelnikov1, A V Kolsanov1, L T Volova1
1Samara State Medical University, Ministry of Healthcare of the Russian Federation, Samara, Russia.
Aim:
To evaluate the biocompatibility of additive materials for personified endoprostheses of hand joints in vivo.
Material And Methods:
We tested a material based on titanium that was implanted into muscles and bone tissue in experiment on rabbits. Follow-up was 30 and 90 days.
Results:
Implantation into muscle tissue is accompanied by reaction against foreign body followed by fibrosis without concomitant inflammation. Induction of osteogenesis and trabecular structures remodeling were detected after implantation into bone tissue.
Conclusion:
Biocompatibility of tested titanium-based material was confirmed.