Related Experiment Video
Updated: Feb 22, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Determination of methotrexate in spiked human blood serum using multi-frequency electrochemical immittance
Sereilakhena Phal1, Britta Lindholm-Sethson1, Paul Geladi2
1Department of Chemistry, Umeå University, SE 90187 Umeå, Sweden.
Abstract:
This article describes an attempt to develop a sensor based on multi-frequency immittance spectroscopy for the determination of methotrexate (MTX) in blood serum using gold electrodes modified with antibodies. The attachment of antibodies was monitored with electrochemical immittance spectroscopy (EIS) and X-ray photoelectron spectroscopy (XPS). The EIS measurements of MTX resulted in a data matrix of size 39 × 55. The data were analysed using multivariate data analysis and showed a concentration dependence and time dependence that could be separated. This allowed the calculation of a multivariate calibration model. The model showed good linear behavior on a logarithmic scale offering a detection limit of 5 × 10-12 mol L-1.