Failure to Treat Life-Threatening Ventricular Tachyarrhythmias in Contemporary Implantable
Anna Margrethe Thøgersen1, Jacob Moesgaard Larsen1, Jens Brock Johansen1
1From the Department of Cardiology, Aalborg University Hospital, Denmark; Department of Cardiology, Odense University Hospital, Denmark (J.B.J.); Paul L. Foster School of Medicine, Texas Tech University Health Science Center, El Paso (M.A.); and Cedars-Sinai Heart Institute, Cedars-Sinai Medical Center, Los Angeles, CA (C.D.S.).
Background:
In clinical trials, manufacturer-specific, strategic programming of implantable cardioverter-defibrillators (ICDs), including faster detection rates, reduces unnecessary therapy but permits therapy for ventricular tachycardia/ventricular fibrillation (VF). Present consensus recommends a generic rate threshold between 185 and 200 beats per minute, which exceeds the rate tested in clinical trials for some manufacturers. In a case series, we sought to determine the relationship between programmed parameters and failure of modern ICDs to treat VF.
Methods And Results:
We reviewed cases in which normally functioning ICDs failed to deliver timely therapy for VF from April 2015 to January 2017 at 4 institutions. Of 10 ambulatory patients, 5 died from untreated VF, 4 had cardiac arrests requiring external shocks, and 1 was rescued by a delayed ICD shock. VF did not satisfy programmed detection criteria in 9 patients (90%). Seven of these patients had slowest detection rates that were consistent with generic recommendations but not tested in a peer-reviewed trial for their manufacturer's ICDs. Manufacturer-specific factors interacted with fast detection rates to withhold therapy, including strict VF episode termination rules, enhancements to minimize T-wave oversensing, and features that restrict therapy to regular rhythms in ventricular tachycardia zones. Untreated VF despite recommended programming accounted for 56% of sudden deaths and 11% of all deaths during the study period.
Conclusions:
Complex and unanticipated interactions between manufacturer-specific features and generic programming can prevent therapy for VF. More data are needed to assess the risks and benefits of translating evidence-based detection parameters from one manufacturer to another.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Dysrhythmias VI: Management of Dysrhythmias
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiomyopathy II: Dilated Cardiomyopathy
Antiarrhythmic Drugs: Class III Agents as Potassium Channel Blockers