Related Experiment Video
Updated: Feb 22, 2026

Clinical Imaging of Microwave Mammography
Published on: November 14, 2025
A novel 3-dimensional electromagnetic guidance system increases intraoperative microwave antenna placement accuracy
Amit V Sastry1, Jacob H Swet1, Keith J Murphy1
1Division of Hepato-Pancreato-Biliary Surgery, Department of Surgery, Carolinas Medical Center, Charlotte, NC, USA.
Background:
Failure to locate lesions and accurately place microwave antennas can lead to incomplete tumor ablation. The Emprint™ SX Ablation Platform employs real-time 3D-electromagnetic spatial antenna tracking to generate intraoperative laparoscopic antenna guidance. We sought to determine whether Emprint™ SX affected time/accuracy of antenna-placement in a laparoscopic training model.
Methods:
Targets (7-10 mm) were set in agar within a laparoscopic training device. Novices (no surgical experience), intermediates (surgical residents), and experts (HPB-surgeons) were asked to locate and hit targets using a MWA antenna (10-ultrasound only, 10-Emprint™ SX). Time to locate target, number of attempts to hit the target, first-time hit rate, and time from initiating antenna advance to hitting the target were measured.
Results:
Participants located 100% of targets using ultrasound, with experts taking significantly less time than novices and intermediates. Using ultrasound only, successful hit-rates were 70% for novices and 90% for intermediates and experts. Using Emprint™ SX, successful hit rates for all 3-groups were 100%, with significantly increased first-time hit-rates and reduced time required to hit targets compared to ultrasound only.
Discussion:
Emprint™ SX significantly improved accuracy and speed of antenna-placement independent of experience, and was particularly beneficial for novice users.
More Related Videos
09:13Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
10:25Technical Approach for Infrared Tracking for Soft Tissue Navigation with a Holographic Head-Mounted Display and Preclinical Validation