In vitro evaluation of flow patterns and turbulent kinetic energy in trans-catheter aortic valve prostheses
Related Experiment Video
Updated: Feb 22, 2026

Particle Image Velocimetry Investigation of Hemodynamics via Aortic Phantom
Published on: February 25, 2022
Daniel Giese1, Kilian Weiss2,3, Bettina Baeßler2
1Department of Radiology, University Hospital of Cologne, Cologne, Germany. daniel.giese@uk-koeln.de.
Objectives:
The objective of the current work was to evaluate flow and turbulent kinetic energy in different transcatheter aortic valve implants using highly undersampled time-resolved multi-point 3-directional phase-contrast measurements (4D Flow MRI) in an in vitro setup.
Materials And Methods:
A pulsatile flow setup was used with a compliant tubing mimicking a stiff left ventricular outflow tract and ascending aorta. Five different implants were measured using a highly undersampled multi-point 4D Flow MRI sequence. Velocities and turbulent kinetic energy values were analysed and compared.
Results:
Strong variations of turbulent kinetic energy distributions between the valves were observed. Maximum turbulent kinetic energy values ranged from 100 to over 500 J/m3 while through-plane velocities were similar between all valves.
Conclusion:
Highly accelerated 4D Flow MRI for the measurement of velocities and turbulent kinetic energy values allowed for the assessment of hemodynamic parameters in five different implant models. The presented setup, measurement protocol and analysis methods provides an efficient approach to compare different valve implants and could aid future novel valve designs.