Related Experiment Video
Updated: Feb 22, 2026

Intrathecal Application of a Fluorescent Dye for the Identification of Cerebrospinal Fluid Leaks in Cochlear Malformation
Published on: February 29, 2020
Comparison of sprotte and Quincke needles with respect to spinal fluid leakage using artificial spinal cord
Yoshikiyo Amaki1, Michihiko Moriyama1, Toshimichi Kuzuta1
1Department of Anesthesiology, Jikei University School of Medicine, 3-25-8 Nishisinbashi, Minato-ku, 105, Tokyo, Japan.
Purpose:
This research investigated whether the Sprotte needle causes less leakage of CSF than the Quincke needle in the artificial spinal cord.
Methods:
The changes in intradural pressure, extradural pressure, and leaked volume of CSF were evaluated following puncture with Sprotte and Quincke needles in the artificial spinal cord.
Results:
The decrease in intradural pressure was 9.7±1.8 mm H2O with the Sprotte needle and 20.5±2.7 mm H2O with the Quincke needle (P<0.05). The volume of leakage of artificial CSF was 2.0±0.3 ml with the Sprotte needle and 3.3 ±0.3 ml with the Quincke needle (P<0.01). The extradural pressure increase was 166.1±8.2 mm H2O with the Sprotte needle and 186.8±13.2 mm H2O with the Quincke needle (P<0.05).
Conclusion:
The Sprotte needle produces less CSF leakage than the Quincke needle.