Evaluation of the Interference of Hemoglobin Variant J-Bangkok on Glycated Hemoglobin (HbA1c) Measurement by Five

Dong-Mei Wen1, Sheng-Nan Xu1, Wei-Jia Wang1

  • 1Laboratory Medicine Centre, Zhongshan Hospital Affiliated to Sun Yat-sen University, Zhongshan, China.

Objective The interference of the hemoglobin variant (Hb J-Bangkok) was evaluated on 4 different glycated hemoglobin assays and compared with a reference immuno assay. Methods An overall test of coincidence of 2 least-squares linear regression lines was performed to determine whether the presence of Hb J-Bangkok caused a statistically significant difference in HbA1c results compared with a reference immuno assay. Statistical analysis was performed on the difference of the estimated average glucose calculated from HbA1c values and fasting plasma glucose in the Hb J-Bangkok variant group using the different detection systems. Deming regression analysis was used to determinate whether Hb J-Bangkok had a significant interference on HbA1c results using an HbA1c±10% relative bias at 6% and 9% HbA1c as evaluation limits. Results Turbidimetric inhibition immunoassay method, and enzymatic methods were not affected by Hb J-Bangkok. However, Hb J-Bangkok showed statistically significant interference to the two ion-exchange high-performance liquid chromatography methods. Conclusion When performing HbA1c tests, clinical laboratory personnel should identify the Hb variant and select the appropriate methods or use alternative indicators.

Related Concept Videos

Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
1.8K
Assessment of blood pressure in brachial artery(one-step method)01:15

Assessment of blood pressure in brachial artery(one-step method)

This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
1.2K
Blood Typing01:10

Blood Typing

Understanding an individual's blood group is a critical component of transfusion medicine. It ensures compatibility in blood transfusions, organ transplants, and even during pregnancy. Determining these blood groups involves the ABO and Rh blood typing systems, utilizing specific antigens and corresponding anti-sera to identify an individual's blood type.
Antigens are protein molecules that reside on the surface of red blood cells (RBCs). The ABO and Rh blood typing systems target...
3.1K
Errors occurring during blood pressure monitoring01:25

Errors occurring during blood pressure monitoring

Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
1.6K
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.7K