Related Experiment Video
Updated: Aug 11, 2026

Methods for the Discovery of Novel Compounds Modulating a Gamma-Aminobutyric Acid Receptor Type A Neurotransmission
Published on: August 16, 2018
[Screening for tetracyclic benzodiazepines using EMIT-st (benzodiazepines) and TDx]
1Institut für Rechtsmedizin der Universität Giessen, Bundesrepublik Deutschland.
Abstract:
The article reports threshold ranges and detection limits for the screening of tetracyclic benzodiazepines (adinazolam, alprazolam, brotizolam, estazolam, loprazolam, midazolam, triazolam and many metabolites) using the EMIT-st and TDx system (FPIA). In most cases, the cross reactivities of the two systems are comparable, but there are also remarkable differences. Detection limits are mainly in the range between 0.2 and 0.5 mg/l but, in some cases, the limits are also considerably higher. The within-day precision is 1.1% (TDx) and 16.5% (EMIT-st); day-to-day precision is 4.7% (TDx) and 12.6% (EMIT-st), respectively.