Radiation dose reduction in plain radiography of the full-length lower extremity and full spine
Mi Ran Jeon1, Hee Jin Park1, So Yeon Lee1
1Department of Radiology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Objective:
To compare the diagnostic performance of standard- and low-dose radiographs of the full-length lower extremity and spine.
Methods:
This study included 223 patients who visited our hospital and received full-length lower extremity standing radiographs and full-spine radiographs. We determined the dose area product (DAP) of each image, and effective doses (ED, mSv) were calculated based on the DAP. Subjective evaluation of the full-length radiographs was based on image quality, which was assessed by bony cortex and trabecula evaluation, and on diagnostic performance, which was assessed by leg length measurement. Subjective evaluation of the full-spine radiographs was based on image quality, which was assessed by viewing the vertebral endplate, pedicle and lateral border of vertebral body, and on diagnostic performance from measurement of Cobb's angle.
Results:
For the full-length view and the full-spine view both the mean DAP and ED values of the standard-dose group were significantly higher than those of the low-dose group (p < 0.05). Mean scores for subjective values did not significantly differ based on the radiation dosage (p-values, 0.15-0.99). The subjective value scores for the full-length view were 2.94-2.98 in the standard-dose group and 2.91-3.00 in the low-dose group. Of note, both groups had very high scores. Additionally, the diagnostic performance scores between the two groups were also very high (range from 2.92 to 3.00).
Conclusion:
Reducing mAs by 50% of the standard dose does not affect the radiograph image quality or its clinical validity. Advances in knowledge: Radiation dose reduction (50% of the standard dose of mAs) in plain radiography of the full-length lower extremity and full spine do not affect the clinical validity and the image quality.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Biological Effects of Radiation