Radiation dose and magnification in pelvic X-ray: EOS™ imaging system versus plain radiographs
P Chiron1, L Demoulin1, K Wytrykowski1
1Département d'orthopédie traumatologie, hôpital Pierre-Paul-Riquet, CHU de Toulouse, place du Docteur-Baylac, TSA 40031, 31059 Toulouse cedex 9, France.
Background:
In plain pelvic X-ray, magnification makes measurement unreliable. The EOS™ (EOS Imaging, Paris France) imaging system is reputed to reproduce patient anatomy exactly, with a lower radiation dose. This, however, has not been assessed according to patient weight, although both magnification and irradiation are known to vary with weight. We therefore conducted a prospective comparative study, to compare: (1) image magnification and (2) radiation dose between the EOS imaging system and plain X-ray.
Hypothesis:
The EOS imaging system reproduces patient anatomy exactly, regardless of weight, unlike plain X-ray.
Material And Method:
A single-center comparative study of plain pelvic X-ray and 2D EOS radiography was performed in 183 patients: 186 arthroplasties; 104 male, 81 female; mean age 61.3±13.7years (range, 24-87years). Magnification and radiation dose (dose-area product [DAP]) were compared between the two systems in 186 hips in patients with a mean body-mass index (BMI) of 27.1±5.3kg/m2 (range, 17.6-42.3kg/m2), including 7 with morbid obesity.
Results:
Mean magnification was zero using the EOS system, regardless of patient weight, compared to 1.15±0.05 (range, 1-1.32) on plain X-ray (P<10-5). In patients with BMI<25, mean magnification on plain X-ray was 1.15±0.05 (range, 1-1.25) and, in patients with morbid obesity, 1.22±0.06 (range, 1.18-1.32). The mean radiation dose was 8.19±2.63dGy/cm2 (range, 1.77-14.24) with the EOS system, versus 19.38±12.37dGy/cm2 (range, 4.77-81.75) with plain X-ray (P<10-4). For BMI >40, mean radiation dose was 9.36±2.57dGy/cm2 (range, 7.4-14.2) with the EOS system, versus 44.76±22.21 (range, 25.2-81.7) with plain X-ray. Radiation dose increased by 0.20dGy with each extra BMI point for the EOS system, versus 0.74dGy for plain X-ray.
Conclusion:
Magnification did not vary with patient weight using the EOS system, unlike plain X-ray, and radiation dose was 2.5-fold lower.
Level Of Evidence:
3, prospective case-control study.
More Related Videos
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies IV: Magnetic Resonance Imaging
Biological Effects of Radiation