Comparison of radiation doses imparted during 128-, 256-, 384-multislice CT-scanners and cone beam computed
N Guberina1, U Dietrich1, D Arweiler-Harbeck2
1Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany.
Purpose:
To examine radiation-doses imparted during multislice (MSCT) and cone-beam computed-tomography (CBCT) for perioperative examination of cochlear-implant insertion.
Methods:
Radiation-doses were assessed during standardized petrous-bone CT-protocols at different MSCT ((I) single-source CT-scanner Somatom-Definition-AS+, (II) 2nd generation of dual-source CT-scanner Somatom-Definition-Flash, (III) 3rd generation of dual-source CT-scanner Somatom-Force and at the CBCT Ziehm-Vision-RFD3D ((IV) (a) RFD-3D (Standard-modifier), (b) RFD-3D (Low-dose-modifier)). Image quality was examined by two radiologists appraising electrode-array placement, quality-control of cochlear-implant surgery and complications based on real patients' examinations (n=78).
Results:
In MSCT-setting following radiation-doses were assessed (CTDIw; DLP): (I) 21.5mGy; 216mGycm; (II) 19.7mGy; 195mGycm; (III) 12.7mGy; 127mGycm; in the CBCT setting radiation doses were distributed as follows: (IV) (a) 1.9mGy; 19.4mGycm; (b) 1.2mGy; 12.9mGycm. Overall, image quality was evaluated as good for both, MSCT- and CBCT-examinations, with a good interrater reliability (r=0.81).
Conclusion:
CBCT bears considerable dose-saving potential for the perioperative examination of cochlear-implant insertion while maintaining adequate image quality.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography