Radiation dose reduction during neuroendovascular procedures
Pearse P Morris1, Carol P Geer2, Jasmeet Singh2
1Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Aim:
To describe the impact of steps towards reduction of procedural doses of radiation during neuroendovascular procedures.
Methods:
Phantom exposures under controlled circumstances were performed using a Rando-Alderson adult-sized head phantom. Customized imaging protocols were devised for pediatric and adult imaging and implemented in clinical use. Outcome data for estimated skin doses (ESD) and dose-area product (DAP) following pediatric and adult diagnostic and interventional procedures over 4.5 years were analyzed retrospectively.
Results:
Dose estimates were reduced by 50% or more after introduction of customized imaging protocols in association with modification of personnel behavior compared with doses recorded with previously used vendor-recommended protocols.
Discussion:
Substantial reductions in radiation use during neuroendovascular procedures can be achieved through a combination of equipment modification and operator behavior.
Related Concept Videos
Aneurysm IV: Nursing Management
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...