The algorithmic performance of J-Tpeak for drug safety clinical trial
Simon C Chien1, Richard E Gregg2
1Advanced Algorithm Research Center, Philips Healthcare, Andover, MA, USA; Department of Electrical Engineering, California Institute of Technology, Pasadena, CA, USA.
Introduction:
The interval from J-point to T-wave peak (JTp) in ECG is a new biomarker able to identify drugs that prolong the QT interval but have different ion channel effects. If JTp is not prolonged, the prolonged QT may be associated with multi ion channel block that may have low torsade de pointes risk. From the automatic ECG measurement perspective, accurate and repeatable measurement of JTp involves different challenges than QT. We evaluated algorithm performance and JTp challenges using the Philips DXL diagnostic 12/16/18-lead algorithm. Measurement of JTp represents a different use model. Standard use of corrected QT interval is clinical risk assessment on patients with cardiac disease or suspicion of heart disease. Drug safety trials involve a very different population - young healthy subjects - who commonly have J-waves, notches and slurs. Drug effects include difficult and unusual morphology such as flat T-waves, gentle notches, and multiple T-wave peaks.
Methods:
The JTp initiative study provided ECGs collected from 22 young subjects (11 males and females) in randomized testing of dofetilide, quinidine, ranolazine, verapamil and placebo. We compare the JTp intervals between DXL algorithm and the FDA published measurements. The lead wise, vector-magnitude (VM), root-mean-square (RMS) and principal-component-analysis (PCA) representative beats were used to measure JTp and QT intervals. We also implemented four different methods for T peak detection for comparison.
Results:
We found that JTp measurements were closer to the reference for combined leads RMS and PCA than individual leads. Differences in J-point location led to part of the JTp measurement difference because of the high prevalence of J-waves, notches and slurs. Larger differences were noted for drug effect causing multiple distinct T-wave peaks (Tp). The automated algorithm chooses the later peak while the reference was the earlier peak. Choosing among different algorithmic strategies in T peak measurement results in the tradeoff between stability and the accurate detection of calcium or sodium channel block.
Conclusion:
Measurement of JTp has different challenges than QT measurement. JTp measurement accuracy improved with combined leads RMS and PCA over lead II or V5.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:31Biosensor-based High Throughput Biopanning and Bioinformatics Analysis Strategy for the Global Validation of Drug-protein Interactions
Published on: December 1, 2020
Related Concept Videos
Noncompartmental Analysis: Miscellaneous Pharmacokinetic Parameters
One key aspect of the noncompartmental approach is determining a drug's total clearance. This can be done by dividing the drug dose by the area under the concentration-time curve from zero to infinity. The area under the concentration-time curve represents the drug's...
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
Model-Independent Approaches for Pharmacokinetic Data: Noncompartmental Analysis
One important characteristic of noncompartmental analyses is that drug exposure increases proportionally with increasing doses. This...
Analysis of Population Pharmacokinetic Data
Clinical Trials: Overview
Pharmacodynamic Models: Emax Drug–Concentration Effect Model