An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Irregular Implant Design Decreases Periimplant Stress and Strain Under Oblique Loading
Ling He1, Jiwu Zhang, Xiucheng Li
1*Research Assistant, Department of The Second Dental Center, Peking University School and Hospital of Stomatology, Beijing, People's Republic of China. †Postgraduate, Department of School of Mathematics and Physical, University of Science and Technology Beijing, Beijing, People's Republic of China. ‡Research Assistant, Department of School of Mathematics and Physical, University of Science and Technology Beijing, Beijing, People's Republic of China. §Attending Doctor, Department of The Second Dental Center, Peking University School and Hospital of Stomatology, Beijing, People's Republic of China. ¶Professor, Department of The Second Dental Center, Peking University School and Hospital of Stomatology, Beijing, People's Republic of China.
Objectives:
To investigate whether a different implant geometry with the same potential contact surface area (PCSA) affects the principal stress and strains in bone.
Material And Methods:
Three-dimensional finite-element models were created with a single endosseous implant embedded in bone. The irregular (IR) dental root-analog implant and regular (R) cylindrical implant with the same PCSA 350 mm were modeled, keeping the size of the thinnest implant wall 0.8 mm, and the thinnest bone wall 1 mm. The regular or irregular abutments were either 4.5 mm lower than the platform of the implants or 5 mm higher than the platform of the implants, both with the taper 1.44°. A 100 N vertical or 100 N vertical/50 N horizontal occlusal loading was applied. The biomechanical behaviors of periimplant bone were recorded.
Results:
The IR implant design experienced lower periimplant stress and strain under oblique loading than that of R implant design. In the IR implant design, comparable stress in bone, implant, and abutment were found under 100 N vertical loading or 100 N vertical/50 N horizontal loading. In the R implant design, much higher stress in bone, implant, and abutment were found under 100 N vertical/50 N horizontal loading than that under 100 N vertical loading.
Conclusion:
Irregular dental root-analog implant is a biomechanically favorable design principle for decreasing periimplant stress and strain under oblique loading.
More Related Videos
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Related Concept Videos
Stress on an Oblique Plane
Normal Strain under Axial Loading
Eccentric Axial Loading in a Plane of Symmetry
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes....
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Unsymmetric Loading of Thin-Walled Members
The concept of the shear center is crucial in countering the...