Temperature Rise on the Plugger Surface of 2 Commercially Available Gutta-percha Heating Devices
Fotis Dimopoulos1, Konstantinos Dervenis1, Christos Gogos1
1School of Dentistry, Aristotle University of Thessaloniki, Thessaloniki, Greece.
Introduction:
The objective of this study was to examine the temperature rise on the plugger surface of 2 commercially available gutta-percha heating devices: the System B (Kerr Dental, Amersfoort, The Netherlands) and the System B Cordless Pack Unit (Kerr Dental).
Methods:
Temperature changes were recorded by a Thermocouple Data Logger device (Pico Technology Ltd, St Neots, UK) and 2 thermocouples: the first to record the temperature on the plugger surface in an isolated polytetrafluoroethylene system and the second to record the base temperature of the environment. The gutta-percha heating devices studied were System B with F, FM, M, and ML pluggers set at the "use" position, "touch" mode, temperature of 200°C, and a power setting of 10 and the System B Cordless Pack Unit with the FM plugger set at low power. Two variables were extracted from the collected temperature data: the temperature on the plugger surface 10 seconds after activating each gutta-percha heating device (θ10) and the time required to reach 60°C (t60). The differences between the pluggers over those 2 variables were investigated using analysis of variance and the Tukey B test for post hoc comparisons (P < .05).
Results:
The mean θ10 for all pluggers ranged between 73°C and 87°C. The mean t60 for all pluggers ranged between 1.3 and 3.3 seconds. No clinically significant differences between the pluggers were observed.
Conclusions:
The gutta-percha heating devices tested achieve maximum temperatures lower than 94°C and are capable of gutta-percha phase transformation within approximately 4 seconds of activation.
Related Concept Videos
Thermal expansion and Thermal stress: Problem Solving
To solve the problem, first, identify the known and unknown quantities. The initial length (L) of the bridge is 1275 m, the coefficient of linear expansion (α) for steel is 12 x 10-6/°C, and the change in temperature (ΔT) is 55...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...