Abstract:
Centrifugation-based whole blood (WB) separation represents the worldwide standard but it depends on electricity and infrastructure. We have prospectively evaluated a novel hollow-fibre WB separation system that does not require manual priming or blood flow regulation (n = 29). RBC units contained sufficient Hb (50·4 g ± 4·3), low leucocytes (90 000 ± 0·008), exhibited low haemolysis (0·57% ± 0·49) and robust ATP content (51·47% ± 8·2) after 43 days storage. Plasma units contained low leucocytes and mean coagulation factor activities for FV, FVIII and FXI were 47%, 90% and 68%, respectively. RBC met quality specifications but plasma units exhibited reduced FV and FXI activity.