Twiddlers syndrome presenting as life threatening electrical storm
Shahrukh Hashmani1, Aamir Hameed Khan2
1Cardiology Resident,Aga Khan University Hospital, Karachi..
A patient experienced inappropriate implantable cardioverter-defibrillator (ICD) shocks due to lead displacement, a complication known as Twiddler's syndrome. This condition required lead repositioning to prevent potentially lethal consequences.
Area of Science:
- Cardiology
- Medical Device Technology
- Electrophysiology
Background:
- Implantable cardioverter-defibrillators (ICDs) are crucial for preventing sudden cardiac death in patients with severe cardiomyopathy.
- Lead integrity and proper positioning are essential for accurate signal sensing and effective defibrillation therapy.
Observation:
- A 75-year-old male with non-ischaemic cardiomyopathy developed inappropriate ICD shocks 13 months post-implantation.
- Over-sensing of the right ventricular lead and double counting of right atrial signals were identified as the cause.
- Chest X-ray revealed significant lead displacement: the right ventricular lead migrated to the right atrium with coiling in the pocket, and the shocking coil was near the tricuspid annulus.
Findings:
- The patient's symptoms were attributed to Twiddler's syndrome, a rare complication involving ICD lead dislodgement.
- Diagnosis was confirmed via chest X-ray, demonstrating the characteristic lead migration pattern.
- Successful management involved removal and replacement of the dislodged ICD lead.
Implications:
- Twiddler's syndrome can lead to inappropriate shocks and device malfunction, posing a significant risk to patient safety.
- This case highlights the importance of vigilant post-implantation monitoring and awareness of lead complication syndromes.
- Prompt recognition and intervention are critical for mitigating the potentially lethal consequences of ICD lead displacement.
More Related Videos
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
Series R—L Circuit Transients
Using Kirchhoff's Voltage Law (KVL) to analyze this circuit helps determine the total asymmetrical fault current, which consists...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias